Intravenous (IV) fluid therapy requires monitoring of the remaining container volume and the infusion condition. Periodic visual inspection can be difficult in busy care environments. This paper presents a designoriented IoT architecture for estimating remaining IV fluid and infusion rate using two sensing paths: a load cell interfaced through an HX711 converter to estimate fluid mass, and an infrared (IR) sensor positioned at the drip chamber to count drops. An ESP32-class controller is proposed for local acquisition, filtering, plausibility checking, time-to-empty (TTE) estimation and wireless status transmission. The load-cell path estimates volume from calibrated mass, while the optical path estimates flow from the observed drop count and the administration-set drop factor. The two estimates are compared to identify inconsistent measurements and possible sensor faults. A configurable multi-level alert model and a local alarm are specified; remote dashboard communication is supplementary and is not required for local alarm operation. An optional tube-clamping actuator is treated as a separate, non-clinical research concept. To avoid presenting an unverified intervention as safe, the initial evaluation is recommended to use alertonly operation, with the actuator physically disabled until a formal hazard analysis and independent validation have been completed. The manuscript defines the mathematical model, calibration procedure, proposed hardware interfaces, test protocol and performance measures. No measured prototype results are claimed in this design paper. Experimental validation using controlled laboratory trials is required before performance or clinical benefit can be established.
Internet of Things, Intravenous infusion monitoring, load cell, HX711, infrared drop sensor, ESP32, multisensor comparison, time-to-empty estimation, predictive alert, safety engineering.